6 Ch 8 kV Common Floating Return HV Board Coming Soon 
Tentative Data
  • 6 channels in a double width (10TE wide) board
  • 8 kV / 3mA output ranges
  • Max output power:
    • 9 W (<3 kV output)
    • 8 W (>3 kV output)
  • Channels with individually selectable positive or negative polarity
  • SHV coaxial output connectors
  • Common floating return
  • Max Ripple smaller than < 30mVpp @ 6kV/1mA or 8kV/800µA (100MHz band width)
  • 200 mV Vset resolution
  • 50 nA Iset resolution
  • Under/over-voltage alert, overcurrent and max. voltage protection
  • Programmable ramp-up / ramp-down (1-500 V/s)
  • Interlock logic for board enable
Power Supplies
SY1527 / SY2527 / SY3527 Universal Multichannel System
SY1527 - SY2527 - SY3527 HV Floating Power Supply Boards
Evaluation of contents:
What do you think about the page contents usefulness?

1 2 3 4 5
mediocre excellent
Overview

The Mod. A1580 is a double width (10TE wide) board that houses 6 independent High Voltage channels. Each channel can provide a ±8kV max voltage, a 3 mA max current and a 9 W max power (8 W max power when output voltage is larger than ±3 kV).
Channels have common floating return (common return insulated from the crate ground); HV outputs are delivered through SHV connectors.
The HV output RAMP-UP and RAMP-DOWN rates may be selected independently for each channel in the range 1÷500 V/s in 1 V/s steps.
Safety features include:

• OVERVOLTAGE and UNDERVOLTAGE warning when the output voltage differs from
the programmed value by more than 2% of set value (minimum 10V).

• Programmable VMAX protection limit

• OVERCURRENT detection: if a channel tries to draw a current larger than its programmed limit, it enters TRIP status, keeping the maximum allowed value for a programmable time (TRIP), before being switched off

• Channels can be enabled or disabled through the Interlock logic.